Karl DEUTSCH Reference Magnet and Zero Field Chamber, Accessory for the field strength meter deutrometer 3873

The application range

Reference magnet and zero field chamber are used for checking and for zero adjustment of the field intensity meter DEUTROMETER 3873 and probes.

 

Function

Reference magnets generate static magnetic fields of known magnitude with which the scale accuracy of the field intensity meter can be checked.

Zero field chambers screen against external magnetic interference fields to a large extent.

With every switch-on the DEUTROMETER carries out an automatic zero adjustment. During this time the probe must not be in any magnetic field.

Thus it is recommended to place the probe in the zero field chamber during switch-on. Also the natural magnetic earth field is shielded to a large extent.

 

Technical data

Reference magnet:

Magnetic field generated in air is (in a distance of 500 mm): H < 0.4 kA/m and B < 0.5 mT.
The individual value is engraved on the case surface..

For more information visit Karl Deutsch website.

Jus taip pat gali dominti

Karl DEUTSCH UV LED Large Area Lamp 3846 for the fluorescent penetrant and magnetic particle testing

The multifunctional UV LED large area lamps are ideally suited…

Išsamiau

Karl DEUTSCH UV LED Combi Hand Lamp 3815

The portable hand lamp for mains and battery operation A…

Išsamiau

Karl DEUTSCH Magnetic Particles Testing DEUTROPULS Hand Yokes

Mobile Crack detection: Safe and reliable The DEUTROPULS Hand Yoke…

Išsamiau

Karl DEUTSCH Magnetic Particle Test Units DEUTROPULS

Portable high current generators In different power classes a mobile…

Išsamiau

Karl DEUTSCH FLUXA® Inspection Media for magnetic particle testing

The premium choice for testing under UV-light FLUXA® Ready Concentrates: Easiest…

Išsamiau

Karl DEUTSCH Field Strength Meter DEUTROMETER 3873

Convenient and practice-oriented measurement of field strength The DEUTROMETER 3873…

Išsamiau

Karl DEUTSCH ESV Demagnetizing Coils, Accessory for Magnetic Particle Crack Detection

For the demagnetization of magnetized parts ESV coils are used…

Išsamiau